Outline

2D Luminance Colorimeter UA-200 is a new product of UA series to measure the uniformity of luminance and chromaticity of FPDs and Interior panels in automobile. The UA-200 achieves high chromaticity accuracy on the order of milli-magnitude.
New six functions are added to current application program, which improve UA-200’s usability for Interior panels in automobile, FPD, and Illumination fields.
Standard lens, Wide lens and Telescopic lens are available.
表示画面s.jpg


Lens angle Model name
Standard lens UA-200S
Standard lens with
wide attachment lens
UA-200WS
Telescopic lens UA-200T

Features


Features

• High chromaticity accuracy : New developed optical filter make it possible to achieve high accuracy of within ±0.008.
• Measuring time : one sec : Optimized algorithm allows reducing measuring time. *Luminance measuring only.
• 1.22 mega pixel CCD : 1,280×960 resolution
• Expansion of function : Available for high brightness version 10-50,000cd/m2
• 3 type lens variation : Standard, Wide, Telescopic lens


Applications

° Evaluation of luminance and chromaticity of LCD, its related materials, and OLED.
° Evaluation of light distribution of Interior panel in automobile and Car Audio.
° Evaluation of luminance and color Mura on the LED and OLED illumination.
 

Standard software

○The standard software can operate an instrument and collect and analyze data.
○Two type software are available for your usage.
・Standard Edition : Full function are available, including control of UA-200
・View Edition : Viewing and analyzing are available. Can analyze measured data without an instrument.
○New functions
①Variety view function
・Focusing assist
・Multipoint extraction (Max 999 points)
②Arbitrary shape of measuring area
・Select measuring area from polygon, square, and circle.
③Multi-area correction
・Splitting into 64×64 (Max) area and applying correction factor to each area.
④Extraction and measuring of multi bright section
・Multiple bright sections are detected automatically by preset threshold.
⑤Diagonal Correction
・Correcting tilting image
・Diagonal correction is available for continuous measurement by single recipe
⑥Link to LabVIEW
・Supporting NATIONAL INSTRUMENTS’ development tool  LabVIEW

 
○View mode (Current model)
Pseudo Color View   Contour View   Split Spot
擬似カラーs.jpg   等高線ビューs.jpg   分割スポットs.jpg
The difference in luminance or color on the image is presented by using colored image.
Suited for Mura evaluation
  Indicates the luminance and color difference by using contour line.   Image is dividend with grid pattern and show value of luminance and chromaticity in each gird.

Standard Spot View   Random Spot View   Cross Section View
規格スポットs.jpg   任意スポットs.jpg   断面図ビューs.jpg
Four type of measurement according to JEITA standard (EIAJ ED-2522/ED-2710). shape and quantity of spots are changeable.   999 points at max can be placed on the image.   Measured data on the cross-section line is shown as graph. The cross-section line can be selected from cross line and diagonal line.

Color diagram View(x,y/u',v')   3D View   Histogram View
色度図ビューs.jpg   3Dビューs.jpg   ヒストグラムs.jpg
Measured data on each view can be plotted on the color diagram   Measured data show as 3D image.   Statistical graph.
Vertical axis means the frequency, horizontal axis means measured data.

Thumbnail View   Time-series graph view    
サムネイルビューs.jpg   時系列ビューs.jpg    
Measured images show as thumbnails.   Variation of measured data with lapse of time shows    

Measured data on each view can be saved as csv format.
Image on each View can be saved as BMP/JPG/PNG file.
 

Specification


Specifications


Model type
UA-200
Photo detector
1.3 million-pixel CCD
Objective lens
Standard lens:f=8.0mm 、 Wide lens:f=5mm
Telescopic lens : f=35mm
Effective pixels
1280×960
Data bit
14bit
Luminance measurement
range
0.05~1,000,000cd/m2
Luminance
linearity *1,*2
±2%  (1cd/m2 over)
±3% (0.05~1cd/m2)
Chromaticity
accuracy 1 *1,*2
±0.003  (1cd/m2 over)
±0.005  (0.05~1cd/m2)
Chromaticty
accuracy 2 *2,*6
±0.008
In-palen
unevenness
uniformity *1,*3
Luminance:±2%
Chromaticity:±0.003
Repeat
characteristic
Luminance:0.3% (1cd/m2 over)*5、0.5% (0.05~1cd/m2)*5
Luminance:0.001  (1cd/m2 over)*4
Chromaticity:0.002  (0.05~1cd/m2 )*4
Measurement time
Luminance:Fastest about 1 sec  100cd/m2
(include communication time)
Luminance & Chromaticity : Fastest about 3 sec 100cd/m2
(include communication time)
 Stability *1,*2
Luminance:1%
Repeatability *1,*2
Luminance:2%
Tempreture
characteristic
Luminance:±3% (Tempreature : 0~40℃)
Humidity characteristic
Luminance:±3% (Humidity: 85%R.H. and below、No condensation)
Interface
LAN (Gigabit Ethernet)
Power supply AC 100V~240V (50/60Hz) / Dedicated AC adapter
Use conditions Tempreature:0℃~40℃
Humidity:85%R.H. and below、No condensation
Outer dimensions (W×D×H)
280×162×194mm(UA-200S)
311×162×194mm(UA-200WS)
307×162×194mm(UA-200T)
Weight
3.9kg (UA-200S)
4.12kg (UA-200WS)
4.45kg (UA-200T)

*1 : Standard sources A
*2 :Center of the CCD
*3 :evaluated on the center of the CCD
*4 :Max - Min
*5 from 10 times continuous measurement
*6 : with combination of standard illuminant and color glass filter
       (O-55,'-48,A-73B,IRA-05,T-44,R-61,B-46,V-44,G-54)

 

Measurement field

UA-200S:Standard lens
Measurement distance
(mm)
300 500 1000 1500 2000 2500
inch 8.9 14.2 27.6 40.9 54.1 67.5
Horizontal(mm) 181.5 289.3 561.0 830.8 1099.2 1371.1
Vertical(mm) 136.1 217.0 420.8 623.1 824.4 1028.6












UA-200WS:Standard lens with wide attachment lens

Measurement
distance
(mm)
300 500 1000 1500 2000 2500
inch 14.5 23.0 44.2 65.3 86.6 107.9
Horizontal(mm) 293.9 466.4 898.7 1326.8 1759.0 2192.9
Vertical(mm) 220.4 349.8 674.0 995.1 1319.2 1644.7











UA-200T:Telescopic lens

Measurement
distance
(mm)
400 500 1000 1500 2000 2500
inch 4.7 5.3 8.3 11.2 14.3 17.3
Horizontal(mm) 95.8 108.2 168.2 228.1 289.9 351.6
Vertical(mm) 71.9 81.1 126.2 171.1 217.4 263.7










*The measurement distance is the distance from the tip of the metal fixture on the instrument of the objective lens.

Operating condition

 OS
Windows XP Professional Service Pack 2 以上(32bit)
Windows 7 Ultimate(32bit/64bit)
Windows 7 Professional(32bit/64bit)
※Not available for Windows XP of 64bit
 CPU
Intel Core i5 (4 Core 2.8GHz) or more
 メモリ
4GB or more
HDD 1GB or more
 Port
Gigabit Ethernet :1 port
*when conducting color correction, the UA-200 require RS-232C serial or USB 1.1
Display 1024×768 or more
1677 million color (32bit) or more
 Others
CD-ROM drive


Product configuration

  • UA-200 main body
  • AC adapter
  • LAN cable
  • Apprication CD-ROM (Apprication software/Instructions manual/SDK file)
  • Lens cap
  • Carrying case 

Available Accessories

  • Wide attachment lens :AL-UA3(for UA-200S)
  • Tripod :Tripod 5N
  • Fine control table   :S-4