Near Infrared Spectroradiometer SR-NIR
High-accuracy measuring with spectrometry for near-infrared region 600-1030nm.
Outline
Recently, demands for measuring near-infrared spectral region have increased, however, especially in
FPD and lighting market, there is not handy and high accuracy measuring instrument.
Under these circumstances, we have developed NIR Spectroradiometer SR-NIR based on technology of
our famous SR-3AR.
And, you can measure Spectral radiance from 380nm to 1030nm using by both other SR series and SR-NIR.
Features
■ Measurable ultra weak optical signal of near-infrared spectral region from flat panel display
■ High accuracy measuring of near-infrared spectral region.
■ Measurable near-infrared spectral form 380nm to 1030nm using by both other SR series and SR-NIR.
Principal use
■ For measuring near infra-infrared spectral region from FPD such as LCD and OLED.
■ For measuring Ne, Ar line spectra.
■ For measuring transmission characteristics of film sheet in near-infrared spectral region.
■ Near-infrared spectral region from other light source.
Specifications
Photo detector | Electronically cooled linear CCD |
---|---|
Dispersing element | Diffraction grating |
Optical system | Object lens : f= 82mm F2.5, Eyepiece lens : 5° view field |
Measurement angle | 2°/1°/0.2°/0.1° Selectable (Electric motor drive) |
Minimum measurement area |
2°: 10.0 mmφ 1°: 4.99 mmφ 0.2°: 1.0 mmφ 0.1°: 0.5 mmφ |
Measurement distance | 350mm – ∞ (Distance from metallic tip of objective lens) |
Wavelength range | 600 – 1030nm |
Spectral bandwidth | 6 – 8nm (Half bandwidth) |
Wavelength resolution | 1nm |
Measurement mode | Auto/Manual (Integral time/frequency) |
Measurement object | Spectral radiance : W・sr-1・m-2・nm-1 |
Calculation function | Spectral radiance : 600 – 1030nm |
Measurement range(*1) |
2° : 0.5 – 3,000 cd/m2 1° : 1 – 9,000 cd/m2 0.2° : 20 – 70,000 cd/m2 0.1° : 100 – 300,000 cd/m2 |
Repeatability(*2) | ±2% or less |
Polarization error | Spectral radiance 5% or less |
Calibration standard |
Topcon standard lamp (Standard illuminant A, 23℃±3℃, 65%RH or less) |
Measurement time(*3) | Approx. 1 – 31 sec |
Interface | RS-232C, USB 2.0 |
Power supply | Dedicated AC adapter AC100V-240V, 50/60Hz, |
Power consumption | Approx. 34W |
Use conditions |
Temperature: 5 – 35℃, Humidity: 80%RH or less (No condensation) |
Outer dimensions (W×D×H) | 150×406×239mm |
Weight | Approx. 5.5kg |
*1 This product is no brightness measurement, so this is the reference data for standard illuminant A.
*2 Wavelength is from 600 to 1030nm, For TOPCON standard illuminant.
*3 Except communication time with PC.
Measurement area
Measurement angle |
Measurement distance [mm] * | ||||
---|---|---|---|---|---|
350 | 500 | 800 | 1000 | 2000 | |
2° | 10.0 | 15.1 | 25.4 | 32.2 | 66.4 |
1° | 4.99 | 7.55 | 12.7 | 16.1 | 33.2 |
0.2° | 1.00 | 1.51 | 2.54 | 3.22 | 6.64 |
0.1° | 0.50 | 0.76 | 1.27 | 1.61 | 3.32 |
* The measurement distance is the distance from the tip of the metal fixture on the instrument of the objective lens.
Accessories
■ Attachment Lens : AL-6
■ Attachment Lens : AL-11
■ Attachment Lens : AL-12
■ CCD Adapter : IA-2
■ Fiber probe : FP-3P
■ Tripod 5N : Tripod5N